Looking after your pump starts with separating the washable parts from the electronic motor. The following routine follows sections 4, 5 and 6 of the supplied K8 manual, with the confirmed 5–10-minute boiling time from section 4.2.

1. Separate and rinse immediately after use

Remove all accessories before cleaning and disinfection. Washable components include the dust cover, flange, flange insert, diaphragm, valve and milk collector.

Rinse separated parts that touch the breast or breast milk in potable water at 25–30°C to remove milk residue. Do this immediately after use rather than leaving milk on the parts.

Keep the pump motor out of the wash basin. It contains electronics and must never be immersed in water.

2. Soak, wash and rinse again

Use a clean basin reserved for infant-feeding equipment, mild dish soap and a clean dishcloth or soft brush.

  • Soak washable parts in 40–60°C soapy water for 3–5 minutes.
  • Wash each part with the clean cloth or soft brush.
  • Rinse with clean water until the soap is removed.

A rinse alone is not the complete cleaning routine. Do not put unwashed parts away for the next session.

3. Disinfect only the specified parts

The manual’s boiling list names four parts: milk collector, diaphragm, flange and valve.

Place those parts in a clean pot with enough water to cover them. Boil at 100°C for 5–10 minutes, keeping the parts away from the pot sides. Allow the water to cool, then gently remove the parts with tongs.

Never boil the electronic motor. Do not extend this boiling instruction to the dust cover, flange inserts or other accessories without specific manufacturer confirmation.

These manual sections do not provide instructions for dishwasher, microwave, UV or steam sterilising. Do not assume that a method suitable for another pump is suitable for your K8.

4. Let everything air-dry completely

Place clean parts on a clean surface or towel and allow them to air-dry. Store them in a clean, cool place only when they are fully dry.

Do not store wet or damp parts in the travel case and do not assemble a pump with moisture trapped inside.

5. Wipe the motor separately

Use a clean, damp cloth to clean the motor. For surface disinfection, sections 4.2 and 6 specify a low-level disinfectant such as ethyl alcohol or isopropyl alcohol at 70–90%.

Keep the motor electronics dry. Surface wiping is not an instruction to soak the motor or put it under a tap.

6. Assemble with clean hands and check the seals

Wash your hands with soap and water, inspect the clean pump kit and assemble only when all parts are dry.

  1. Gently push the flange and valve into the milk collector, leaving no gaps.
  2. Fit the silicone diaphragm onto the motor and press around the edge for a complete seal.
  3. Join the pump motor and milk collector, again checking for gaps.
  4. Choose the flange size suited to you, using the ruler inside your K8 pack.

Do not improvise a different assembly order based on a different pump model.

Keep an eye on parts and milk levels

The manual recommends replacing washable parts such as the valve, diaphragm, milk collector, dust cover and flange every 2–3 months, depending on use.

Each collector holds up to 150ml, and the pump automatically stops a session after 20 minutes. Monitor the milk level while pumping: the timer does not measure how full the collector is.

The K8 is a single-user product. Do not rent or share it between mothers, even after sterilising.